# Vietnam Medical Device Classification (A–D) API

> Classify a medical device or IVD into class A, B, C or D by the Vietnamese rules (Circular 05/2022) and get the declaration or registration route.

- Endpoint: `GET | POST https://apibox.io.vn/v1/vn-medical-device-class`
- Category: Data & JSON
- HTML docs: https://apibox.io.vn/docs/vn-medical-device-class/ (Vietnamese: https://apibox.io.vn/vi/docs/vn-medical-device-class/)
- Online tool: https://apibox.io.vn/tools/vn-medical-device-class/
- Authentication: `Authorization: Bearer <API key>` or `X-API-Key: <API key>` (free key: https://apibox.io.vn/dashboard/#signup)

Answer the rule questions: `invasive` (none, orifice, surgical, implant), `duration` or `duration_minutes` (transient < 60 min, short 60 min – 30 days, long > 30 days), `active` (therapeutic, diagnostic, administer, other) and flags such as contact with the heart / central circulation or central nervous system, biological effect or absorption, medicinal substance, animal or microbial origin, disinfection, contraception, dental, life support. For in-vitro diagnostics send `ivd=true` and one or more `ivd_purpose` values (transfusion screening, HIV-type infections, blood grouping, STI, self-test, point-of-care glucose…).

The API applies the 16 non-IVD rules and 7 IVD rules of Circular 05/2022/TT-BYT Appendix I and returns every rule that applies, the decisive one and the highest class (Decree 98/2021 Article 5). It adds the route: class A / B declaration of applicable standards at the provincial Department of Health, class C / D registration number from the Ministry of Health (30-day review, fast track for FDA / TGA / EU / Japan / NMPA approvals), unlimited validity, trading conditions for B–D and the ISO 13485 condition for imports.

`q=` gives a typical class for common devices (mask, syringe, blood-pressure monitor, glucose meter, condom, stent…) with a confidence level. The registrant remains responsible for the classification.

## Parameters

| Name | In | Type | Required | Default | Description |
|---|---|---|---|---|---|
| `invasive` | body|query | string | no |  | `none`, `orifice` (body orifice), `surgical`, `implant`. |
| `duration` | body|query | string | no |  | `transient`, `short`, `long` (or `duration_minutes`). |
| `active` | body|query | string | no |  | `none`, `therapeutic`, `diagnostic`, `administer`, `other`. |
| `heart_or_central_circulation` | body|query | boolean | no |  | Flags: `heart_or_central_circulation`, `cns_contact`, `biological_effect_or_absorbed`, `delivers_medicine`, `chemical_change_in_body`, `ionizing_radiation`, `dental`, `life_support`, `breast_implant`, `medicinal_substance`, `animal_or_microbial_origin`, `contraception_or_sti`, `xray`, `monitors_vital_dangerous`, `hazardous_energy`, `hazardous_delivery`… |
| `wound` | body|query | string | no |  | Non-invasive wound contact: `barrier`, `dermis`, `dermis_secondary`; also `channel_or_store`, `blood_bag`, `modifies_blood`, `disinfects`. |
| `ivd` | body|query | boolean | no |  | In-vitro diagnostic device; with `ivd_purpose` (comma-separated, e.g. `self_test`, `life_threatening_infection`, `abo_grouping`). |
| `q` | query | string | no |  | Common device name for a typical-class suggestion (used when no rule input is sent). |

## Example request

```bash
curl -X POST "https://apibox.io.vn/v1/vn-medical-device-class" \
  -H "Authorization: Bearer YOUR_API_KEY" \
  -H "Content-Type: application/json" \
  -d '{"invasive":"implant","heart_or_central_circulation":true}'
```

## Example response

```json
{
  "class": "D",
  "class_vi": "Loại D",
  "risk_vi": "mức độ rủi ro cao",
  "decisive_rule": {
    "rule": "Quy tắc 8.2",
    "cls": "D",
    "vi": "Tiếp xúc trực tiếp tim, hệ tuần hoàn trung tâm hoặc thần kinh trung ương"
  },
  "rules_applied": [
    {
      "rule": "Quy tắc 8.2",
      "cls": "D",
      "vi": "Tiếp xúc trực tiếp tim, hệ tuần hoàn trung tâm hoặc thần kinh trung ương"
    }
  ],
  "principle": "Highest class of all applicable rules and intended uses (NĐ 98/2021 Điều 5 khoản 2, 3); devices used together are classified separately (khoản 4).",
  "procedure": {
    "route": "registration",
    "vi": "Đăng ký lưu hành (cấp số lưu hành) tại Bộ Y tế; thẩm định trong 30 ngày kể từ khi nhận hồ sơ đầy đủ, hợp lệ; cấp nhanh khi đã được nước tham chiếu cho lưu hành",
    "ref": "NĐ 98/2021 Điều 21 khoản 1 điểm b, Điều 29, 30, 32",
    "trading": "Cơ sở mua bán phải công bố đủ điều kiện mua bán tại Sở Y tế (trừ danh mục được mua bán như hàng hóa thông thường)",
    "validity_vi": "Số lưu hành có giá trị không thời hạn (trừ số lưu hành cấp khẩn cấp)",
    "validity_ref": "NĐ 98/2021 Điều 21 khoản 3",
    "quality_vi": "TTBYT nhập khẩu: cơ sở sản xuất đạt ISO 13485 và được lưu hành tại bất kỳ nước nào; sản xuất trong nước: cơ sở đã công bố đủ điều kiện sản xuất",
    "quality_ref": "NĐ 98/2021 Điều 23 khoản 1",
    "label_vi": "Nhãn theo pháp luật về nhãn hàng hóa, hướng dẫn sử dụng tiếng Việt, thông tin bảo hành",
    "label_ref": "NĐ 98/2021 Điều 22"
  },
  "reviewed": "2026-11-16",
  "disclaimer": "Reference only, not regulatory advice; the registrant classifies the device and is responsible for it (NĐ 98/2021 Điều 5 khoản 6). Check the official text."
}
```

## Use cases

- Pre-classify products before preparing a Vietnamese declaration or registration file
- Flag medical devices and their class in e-commerce or import catalogues
- Train regulatory staff on the A–D rules with rule-by-rule explanations
- Estimate whether a distributor needs a B/C/D trading declaration

## FAQ

### Are the rules the ASEAN rules?

Circular 05/2022 Appendix I follows the ASEAN Medical Device Directive model: 16 rules for general devices and 7 for IVDs, with the highest applicable class winning.

### Did 2026 change classification?

Circular 24/2026/TT-BYT (from 1/7/2026) links risk levels to the Law on Product Quality and amends Article 8 of Circular 05/2022 only; the classification rules are unchanged.

### Does a registration number expire?

No, registration and declaration numbers have unlimited validity except emergency numbers (Decree 98/2021 Article 21).